HP-ZB Operational Verification
Program Listing (Continued)
3590 WAIT 1000
'
3600 OUTPUT
A
;"BS1"
3610 GOSUB 8540
3620 OUTPUT
A
;"BS0 BXI"
3630 GOSUB 8540
3640 OUTPUT
A
;"BX0 BAl"
3650 GOSUB 8540
3660 OUTPUT
A
;"BA0 BZ1"
3670 GOSUB 8540
3680 OUTPUT
A
;"BZ0 IN AU0
3690 CLEAR
3700 DISP "The
5334A/B is in the READ PEAKS Bmode."
3710 DISP
3720 OUTPUT
A
;"IN FN15
"
3730 DISP D$;"shows '0.00 0.00 b'."
3740 DISP
3750 DISP
F$
3760 PAUSE
3770 CLEAR
3780 OUTPUT
A
;"IN AU0"
3790 GOT0 7960
3800
!
3810
!
CHECKPOINT 6
3820
I$=" Trigger Levels
3830
S=6
3840 GOSUB 7810
3850 DISP
E$;"the 5334A/B0
3860 DISP "will be programmed to the
3870 DISP
3880 DISP "-2.34 1.56
L"
3890 DISP
3900 PAUSE
391 0 REMOTE
A
3920 OUTPUT
6
;"TR0 SE0"
3930 OUTPUT
A
;"AT-2.34 BT+1.56
"
3940 OUTPUT
A
;"TRI FN13"
3950 CLEAR
3960 DISP
D$
3970 DISP "reads '-2.34 1.56 L'."
3980 DISP
3990 DISP
F$
4000 PPlUSE
401 0 CLEAR
4020 DISP "Press
RESET/LOCAL."
4030 DISP "Press READ LEVELS 4 times."
4040 DISP
4050 DISP "This sequence allows the triggerlevel control to return to the front
panel.
"
4060 DISP
4070 DISP "Adjust the
A
and B LEVEL knobs for
a
different level reading."
4080 DISP
E5;"check that"
trigger levels below."
HP
5334B
-
Senice
Manual
4-52