Table
4-4.
HP-IB Operational Verification
Program Listing (Continued)
HP
5334B
-
Service
Manual
4-51
3070 I$=" Input Conditioning-Channel A"
3080
S=4
3090 GOSUB 7810
3100 DISP
E5;"verify that"
3110 DISP "the display annunciators SLOPE,"
3120DISP"X10,AC,50Z,FILTERA,and
COMAlight (inthat order) for"
3130 DISP "INPUT
A.
A
tone
will
sound to"
3140 DISP "mark each input change."
3150 DISP
3160 DISP
A$
31 70 PAUSE
31 80 REMOTE
A
3190 OUTPUT
A
;"IN AU0"
3200 WAIT 1000
..
i
3210 OUTPUT
A
;"AS\"
3220 GOSUB 8540
3230 OUTPUT
A
;"AS0 14x1"
3240 GOSUB 8540
3250 OUTPUT
A
;"AX0 AA1"
3260 GOSUB 8540
3270 OUTPUT
A
;"AA0 AZ1"
3280 GOSUB 8540
3290 OUTPUT
A
;"AZ0 FI1"
3300 GOSUB 8540
3310 OUTPUT
A
;"FI0
CO.1"
3320 GOSUB 8540
3330 OUTPUT
A
;"COO
3340 CLEAR
3350 DISP "The
5334A/B is in the READ PEAKS Anode."
3360 DISP
3370 OUTPUT
A
;"IN FN14
3380 DISP
D5;"shows '0.00 0.00 A'."
3390 DISP
3400 DISP
F5
341 0 PAUSE
3420 CLEAR
3430 OUTPUT
A
;"IN AU0"
3440 GOT0 7960
3450
!
3460
!
CHECKPOINT 5
3470
I$=" Input Conditioning-Channel B"
3480
S=5
3490 GOSUB 7810
3500 DISP
E5;"verify that"
3510 DISP "the display annunciators SLOPE,"
3520 DISP
"X10,
AC,
and 50
Z
light (in thatorder) for INPUT B.
A
tone will"
3530 DISP "sound to mark each input change."
3540 DISP
3550 DISP USING
"lt,K,/"
;
A5
3560 PAUSE
3570 REMOTE
A
3580 OUTPUT
A
;"IN AU0
h