HP5334B
UNIVERSAL
COUNTER
343A
FLUKE
-
OUTPUT
-J
1
TS17-L2
Figure
5-3.
DAC
AIB Gain Adjustment Setup
Equipment:
DC Volt. Std
....................................
FLUKE
343A
Procedure:
Set the
DC
Standard as follows:
................................
Output Voltage
+5.000V dc
2
0.001V
Set the HP 5334B as follows:
READ
LEVELS
............................. Voltage Peaks A
.........................
INPUT Impedance
1
MR (50R key LED off)
Connect the DC Standard to the HP
5334B Input A.
Adjust
AIR102 (A GAIN) for a 5334B front panel display of
+
5.00V dc
5
0.02V dc for both the positive and negative peaks.
Set the HP
5334B as follows:
READ LEVELS
.............................
Voltage Peaks
B
.........................
INPUT Impedance
1
MR (500 key LED off)
6.
Connect the DC Standard to the
HP
5334B Input B.
7.
Adjust AIR103 (B GAIN) for a 5334B front panel display of
+
5.00V dc
4
0.02V dc for for both the positive and negative
peaks.
HP
5334B
-
Service Manual
5-6