400 Series
EN - 66
Environmental parameters (AUX)
Parameter Range Resolution
Transduced
signal
Accuracy
Temperature
-20.0 80.0°C
0.1°C
-20 +80mV
(2.0rdg + 2dgt)
-4.0 176.0°F
0.1°F
-4 +176mV
Humidity
0.0 100.0% RH
0.1% RH
0 +100mV
DC voltage
(0.0 999.9mV)
0.1mV
(0.2 999.9mV)
Luminance
0.001 20.00Lux 0.001 0.02Lux 0 +100mV
0.1 2000Lux 0.1 2Lux 0 +100mV
1 20000Lux 0.1 2Lux 0 +100mV
Mains analisys (PWR)
Frequency measurement
Range [Hz] Resolution [Hz] Accuracy
47.0 63.0
0.1
(2.0rdg + 2dgt)
Voltage range: 5.0 265.0
Current range: 0.005 1.2 x FS
Voltage measurement
Range [V] Resolution [V] Accuracy
5.0 265.0
0.1V
(0.5rdg + 2dgt)
Max crest factor: 1,5
Frequency: 47.0 63.0 Hz
Voltage harmonics measurement
Range [V] Resolution [V] Order Accuracy
0.0 265.0
0.1V
2 15 (2.0rdg + 5dgt)
16 49 (5.0rdg + 10dgt)
Frequency fundamental 47.0 63.0 Hz
Current measurement
Range [A] Resolution [A] Accuracy
0.005 1.2 x FS
See Table 2
(1.0rdg + 2dgt)
Max crest factor: 3
Frequency: 47.0 63.0 Hz
Current harmonics measurement
Range [V] Resolution [A] Order Accuracy
0.005 1.2 x FS
See Table 2
2 15 (2.0rdg + 5dgt)
16 49 (5.0rdg + 10dgt)
Frequency fundamental 47.0 63.0 Hz
Current fundamental 0.020 x FS
Full scale [A] Resolution [A] Full scale [A] Resolution [A]
1 0.001 300 0.1
10 0.01 400 0.1
30 0.01 1000 1
100 0.1 2000 1
200 0.1 3000 1
Table 2: Clamp and instrument full scale and resolution values