First fault current – IT systems
Range [mA] Resolution [mA] Accuracy
0.1 0.9
0.1
(5reading+1digit)
1 999
1
(5reading + 3digits)
Limit contact voltage (ULIM) : 25V, 50V
Test on RCD protection (Molded case type)
Differential protection type (RCD): AC ( ), A ( ), B( ) – General (G), Selective (S) and Delayed ( )
Voltage range Phase-Earth, Phase-Neutral: 100V 265V RCD type AC and A, 190V 265V RCD type B
Rated tripping currents (IN): 10mA, 30mA, 100mA, 300mA, 500mA, 650mA, 1000mA
Frequency: 50/60Hz 5%
Molded case type RCD tripping current - (for General RCD only)
RCD type
I
N Range I
N
[mA]
Resolution [mA] Accuracy
AC, A
I
N = 10mA
(0.3
1.1) I
N
0.1I
N
- 0%, +10%I
N
10mA <IN 650mA
- 0%, +5%I
N
B
30mA IN 100mA
Measurement duration of Molded case type RCD tripping time – TT/TN systems
x 1/2 x 1 x 2 x 5 AUTO
\ G S
G S
G S
G S
G S
G S
10mA
AC 999 999 999 999 999 999 200 250 50 150
310
A 999 999 999 999 999 999 200 250 50 150
310
B
30mA
100mA
AC 999 999 999 999 999 999 200 250 50 150
310
A 999 999 999 999 999 999 200 250 50 150
310
B 999 999 999 999 999 999 310
300mA
AC 999 999 999 999 999 999 200 250 50 150
310
A 999 999 999 999 999 999 200 250 50 150
310
B 999 999 999 999 999 999
500mA
650mA
AC 999 999 999 999 999 999 200 250 50 150
310
A 999 999 999 999 999 999 200 250 310
B
1000mA
AC 999 999 999 999 999 999 200 250
A 999 999 999 999 999 999
B
Table with duration of tripping time measurement [ms] - Resolution: 1ms, Accuracy:(2.0reading + 2digits)
Measurement duration of Molded case type RCD tripping time – IT systems
x 1/2 x 1 x 2 x 5 AUTO
\ G S
G S
G S
G S
G S
G S
10mA
AC 999 999 999 999 999 999 200 250 50 150
310
A
B
30mA
100mA
300mA
AC 999 999 999 999 999 999 200 250 50 150
310
A
B
500mA
650mA
AC 999 999 999 999 999 999 200 250 50 150
310
A
B
1000mA
AC 999 999 999 999 999 999 200 250
A
B
Table with duration of tripping time measurement [ms] - Resolution: 1ms, Accuracy:(2.0reading + 2digits)