22 Excelon XD/XQ Operators Manual
If the initialization process fails, check the following.
- Check drill fixing sponge removed(XD)
- Check carriage fixing sponge removed (Right and Left)
- Check carriage fixing bracket removed
- Check feeler fixing tape removed
If the touch screen does not work, run the touch calibration function.
- Turn off the system and then turn on the system with the [WINDOW] key
pressed.
- After a while touch calibration screen displayed.
- Follow the instruction displayed on the screen.
If the initialization error occurs continuously, follow the instruction below.
- Memorize error message displayed on the screen.
- Turn off the system and then turn on the system with the [STOP] key pushed.
- After a while main screen displayed on the screen without initialization.
- Run Tool Test Mode in the menu bar and check the system.
If the edger work correctly, make sure to reset system time. (see chapter [7.1]
for setup time)
Make sure to turn off the system before insert or remove SD memory card.
If the SD memory card is inserted or removed while edger is running, the data in
the SD card can be corrupted and that data corruption can cause machine
malfunction.
!