N OTE: SureTrend software is required for sections 3.1 to 3.3. For
fur t her details, r efer to t he separat e SureTrend User's Manual locat ed
on t he SureTrend software CD that is packaged with t he lum inom eter.
3 .1
Use r I D
The lum inom et er can stor e a list of 201 User IDs ( User : 0 t o 200) .
Each t est result will be t agged wit h t he current ly select ed User I D.
Select t he Use r I D m enu option, t hen use the and buttons t o
scroll t hrough the list of User IDs, t hen press the but t on t o select
a User I D, or the but t on to cancel t he select ion.
TI P: Use r : 0 is the default select ion if no User I D list has been
defined.
3 .2
Set t i n g u p Pr og r a m Te st Po i n t s a n d Pass/ Fa i l Li m i t s
The lum inom et er has the ability to st ore a list of 5001 unique Progr am
test points ( Pr og# 0 to 5000) , each wit h its own upper ( ) and low er
( ) pass/ fail result t hr eshold values.
Program t est point location nam es can only be program m ed int o the
lum inom eter through the SureTr end software. Progr am pass/ fail
thr esholds can either be set- up on the lum inom eter, or set using the
SureTrend soft ware. I f a location nam e is ent ered and t hr eshold set
in SureTr end, and then downloaded t o t he lum inom eter, the t hreshold
values can’t be changed v ia t he lum inom et er. This is a security feat ure
and prev ents inadv ert ent changes t o the t hresholds after being set in
SureTrend.
TI P: Pr o g : 0 has a special function, whereby t he thr eshold v alues
defined for Pr o g : 0 are used as the initial values when setting-up t he
ot her Pr o g t hresholds. Set t ing Pr o g :0 w it h your typical threshold
values will save t im e when setting- up t he ot her Program s. The
lum inom eter com es wit h pass/ fail t hresholds of 10 RLU for pass and
30 RLU for fail. These pass/ f ail thresholds w ere det erm ined based on
a com pr ehensive st udy of ATP, food r esidue, and m icroorganism
sam ples. To receive t he r eport or m ore infor m ation on set t ing up
pass/ fail thresholds cont act Hy giena or visit our Question and Answer
section at w ww .hygiena.com .
To set t he Progr am result thresholds on the lum inom eter, fir st select
the PROGRAM m enu opt ion by pressing . Scroll to “ PROGRAM” and
pr ess . Then use the and but tons to scroll t hrough the list of
Program s, followed by the but t on to select t he Progr am t o be
changed: