Hygiena SystemSURE Plus - 3.1 User ID; 3.2 Setting up Program Test Points and Pass; Fail Limits

Hygiena SystemSURE Plus
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10
INS0228
July
2020
N OTE: SureTrend software is required for sections 3.1 to 3.3. For
fur t her details, r efer to t he separat e SureTrend User's Manual locat ed
on t he SureTrend software CD that is packaged with t he lum inom eter.
3 .1
Use r I D
The lum inom et er can stor e a list of 201 User IDs ( User : 0 t o 200) .
Each t est result will be t agged wit h t he current ly select ed User I D.
Select t he Use r I D m enu option, t hen use the and buttons t o
scroll t hrough the list of User IDs, t hen press the but t on t o select
a User I D, or the but t on to cancel t he select ion.
TI P: Use r : 0 is the default select ion if no User I D list has been
defined.
3 .2
Set t i n g u p Pr og r a m Te st Po i n t s a n d Pass/ Fa i l Li m i t s
The lum inom et er has the ability to st ore a list of 5001 unique Progr am
test points ( Pr og# 0 to 5000) , each wit h its own upper ( ) and low er
( ) pass/ fail result t hr eshold values.
Program t est point location nam es can only be program m ed int o the
lum inom eter through the SureTr end software. Progr am pass/ fail
thr esholds can either be set- up on the lum inom eter, or set using the
SureTrend soft ware. I f a location nam e is ent ered and t hr eshold set
in SureTr end, and then downloaded t o t he lum inom eter, the t hreshold
values cant be changed v ia t he lum inom et er. This is a security feat ure
and prev ents inadv ert ent changes t o the t hresholds after being set in
SureTrend.
TI P: Pr o g : 0 has a special function, whereby t he thr eshold v alues
defined for Pr o g : 0 are used as the initial values when setting-up t he
ot her Pr o g t hresholds. Set t ing Pr o g :0 w it h your typical threshold
values will save t im e when setting- up t he ot her Program s. The
lum inom eter com es wit h pass/ fail t hresholds of 10 RLU for pass and
30 RLU for fail. These pass/ f ail thresholds w ere det erm ined based on
a com pr ehensive st udy of ATP, food r esidue, and m icroorganism
sam ples. To receive t he r eport or m ore infor m ation on set t ing up
pass/ fail thresholds cont act Hy giena or visit our Question and Answer
section at w ww .hygiena.com .
To set t he Progr am result thresholds on the lum inom eter, fir st select
the PROGRAM m enu opt ion by pressing . Scroll to PROGRAM and
pr ess . Then use the and but tons to scroll t hrough the list of
Program s, followed by the but t on to select t he Progr am t o be
changed: