26
INS0228
July
2020
7 .3
Lu m i n om et e r Er r o r Cod e s
During norm al operation, the lum inom et er perform s
various self-check s on its internal com ponent s. I f a
problem is det ect ed, the display will show an error
num ber:
E
rr
o
r
Code
Possible
Causes
E1
Not
applicable
on
the
luminometer
E2
Temperature
out
of
range
The
luminometer
is
being
used
outside
of
the
specified
operating
temperature
range
(see
section
10)
The
luminometer
has
been
stored
in
an
environment
which
is
outside
of
its
specified
operating
temperature
range
Luminometer
damaged
or
faulty
E3
Erratic
measurement
Luminometer
environment
unstable
or
used
in
an
area
of
high
electromagnetic
noise
Luminometer
tilted
while
measurement
being
performed
Luminometer
lid
damaged
and
allowing
light
in
Protective
pocket
dirty
or
severely
scratched
Luminometer
damaged
or
faulty
E4
User
settings
undefined
The
user
configurable
settings
are
undefined
-
check
and
re-set
the
CLOCK
format,
Use
r
:,
Plan:
and
P
r
og:
data
E5
Program,
Test
Plan
or
User
ID
undefined
Program
thresholds
not
defined,
and
have
been
reset
to
the
default
values
(10
and
30 )
Test
Plan
data
not
defined
User
ID
not
defined
E6
Calibration
self-
checks
failed
Power
the
luminometer
off
and
then
back
on
Luminometer
operating
environment
unstable
Protective
pocket
dirty
or
severely
scratched
Lid
not
fully
closed
Lid
seal
damaged
Protective
pocket
damaged
Luminometer
damaged
or
faulty
E7
Internal
memory
failure
Batteries
are
low
or
loose
Luminometer
memory
damaged
or
faulty
E8
Internal
reader
fault
Batteries
are
flat
or
loose
Luminometer
sample
reader
is
damaged
or
faulty
E9
Internal
error
Batteries
are
flat
or
loose
Luminometer
dropped
or
subjected
to
shock
or
vibration
Luminometer
damaged
or
faulty

Related product manuals