NanoPhotometer
®
N120/NP80/N60/N50/C40 User Manual Version 4.3.1
106
Note: A new Blank is recommended when changing the dilution lid.
4. Set Start and End Wavelength to define the scan range.
Note: If samples with a different wavelength range are selected, the graphs are shown on
full scan range of 200900 nm (N50: 200-650 nm).
5. Baseline correction is set off by default. Enabling the baseline corrections shows a list with
different wavelength options: 377 nm, 604 nm, 650 nm, 770 nm (N/A N50) and 823 nm
(N/A N50). Option to enter any wavelength between 200 nm and 900 nm (N50: 650 nm)
6. Option to smooth the graph with different boxcars. Options: Off, 1 = boxcar 11 (default), 2
= boxcar 21 and 3 = boxcar 61
7. Option to set/calculate a dilution factor for manual diluted samples.