It is also possible to use cuvettes with 5 mm, 2 mm, 1 mm or 0.5 mm path lengths, but
there may be an adapter necessary. Please ask your cuvette supplier for a suitable
adapter.
The minimum volume for accurate measurements depends on the cuvette type used; it
is necessary that the light passes through the sample for accurate measurements.
Center height is 8.5 mm.
Note: The cuvette holder is not removable. Do not pour any cleaning solution into the
cuvette holder as larger amounts of liquids can get into the instrument and cause damage.
SOLVENT COMPATIBILITY (N50/N60/N120/NP80/C40 WITH
SUBMICROLITER CELL)
Most solvents typically used in life science laboratories are compatible with the
NanoPhotometer
®
NanoVolume sample surfaces. The following solvents are compatible for
use with the NanoPhotometer
®
models N50, N60, N120 and NP80 at room temperature:
Acetone (≤5%)
Acetonitrile
Benzene
Butanol
Carbon tetrachloride
Chloroform
Ethanol
Ether
HEPES
Hexane
Isopropanol
MES
Methanol
Methylene chloride
MOPS
Phenol (≤1%)
N-propanol
Toluene
Phosphate containing buffers
PBS (pH 4-10)
Citrate
Borate
Chloride salts
Acids pH 2
Bases pH 10
Note: Highly concentrated acids and bases are not recommended. It is recommended to
wipe the sample surface with a lint-free laboratory wipe immediately upon completion of
each measurement. For more information about compatibility of specific solvents not listed
above, please contact the Implen support team (support@implen.de) to check the
compatibility.