Note: A new Blank is recommended when changing the dilution lid.
4. Enter desired wavelengths ( 1-1 and 1-2) for ratio calculation. It is possible to measure
up to 20 absorbance/ratios simultaneously. More wavelengths for ratio calculation can be
added by selecting the Add Ratio button. Added ratios can be deleted with .
5. Baseline correction is set off by default. Enabling the baseline corrections shows a list with
different wavelength options: 377 nm, 604 nm, 650 nm, 770 nm (N/A N50) and 823 nm (N/A
N50). Option to enter any wavelength between 200 nm and 900 nm (N50: 650 nm)
6. Option to smooth the graph with different boxcars. Options: Off, 1 = boxcar 11 (default), 2 =
boxcar 21 and 3 = boxcar 61
7. Option to set/calculate a dilution factor for manual diluted samples.