NanoPhotometer
®
N120/NP80/N60/N50/C40 User Manual Version 4.3.1
115
lids see page 52 Submicroliter Cell Measurements Basics). Required sample volume for
each lid is shown.
Note: A new Blank is recommended when changing the dilution lid.
4. Default wavelength is 260 nm but can be changed in the range of 200-
900 nm (N50: 200650 nm), depending on the sample/application.
5. Enter a factor for concentration calculation.
6. Unit selection
7. Baseline correction is set off by default. Enabling the baseline corrections shows a list with
different wavelength options: 377 nm, 604 nm, 650 nm, 770 nm (N/A N50) and 823 nm
(N/A N50). Option to enter any wavelength between 200 nm and 900 nm (N50: 650 nm)