82
W/o background and with dye correction:
DOL =
A
max, dye
* ε
prot
A
280
cf
dye
* A
max, dye
* ε
dye
W/o background and w/o dye correction:
DOL =
A
max, dye
* ε
prot
A
280
* ε
dye
DOL Degree of labeling/dye per protein ratio
A
max,dye
Absorbance value at the absorbance maximum of the dye (10 mm path)
A
280
Absorbance at 280 nm (10 mm path)
A
BKG
Absorbance at selected background wavelength (10 mm path)
ε
dye
Extinction coefficient of dye in M
-1
*cm-
1
ε
prot
Extinction coefficient/protein factor in g*cm/l
cf
dye
Dye-dependent correction factor at 280 nm
Table 4. Dye Types, Absorbance Max, Extinction coefficients, and dye-dependent
correction factors
NanoPhotometer
®
Models
Dye Type
Absorbance
maximum
of Dye (nm)
Dye-dependent
ext. coeff.
ε
dye
in M
-1
*cm
-1
Dye-dependent
correction
factor at 280 nm
cf
dye
NP80/N60/N50/C40
Alexa Fluor 350
346
19,000
0.19
NP80/N60/N50/C40
Alexa Fluor 405
401
34,500
0.70
NP80/N60/N50/C40
Alexa Fluor 488
494
71,000
0.11
NP80/N60/N50/C40
Alexa Fluor 532
530
81,000
0.09
NP80/N60/N50/C40
Alexa Fluor 546
554
104,000
0.12
NP80/N60/N50/C40
Alexa Fluor 555
555
150,000
0.08
NP80/N60/N50/C40
Alexa Fluor 568
577
91,300
0.46
NP80/N60/N50/C40
Alexa Fluor 594
590
73,000
0.56
NP80/N60/C40
Alexa Fluor 647
650
239,000
0.03
NP80/N60/C40
Alexa Fluor 680
679
184,000
0.05
NP80/N60/C40
Alexa Fluor 790
785
260,000
0.08
NP80/N60/N50/C40
Cy3
550
150,000
0.05
NP80/N60/C40
Cy5
649
250,000
0.05
NP80/N60/C40
DyLight 649
654
250,000
0.04
NP80/N60/N50/C40
DyLight 488
493
70,000
0.15
NP80/N60/N50/C40
FITC
495
68,000
0.30
NP80/N60/N50/C40
Pacific Blue
409
30,000
0.20
NP80/N60/N50/C40
Pacific Orange
397
24,500
0.60
NP80/N60/N50/C40
pHrodo Green
505
75,000
0.20
NP80/N60/N50/C40
pHrodo Red
560
65,000
0.12
NP80/N60/N50/C40
r-PE
566
200,000
0.18
NP80/N60/N50/C40
Texas Red
595
80,000
0.18

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Implen NanoPhotometer NP80 and is the answer not in the manual?

Implen NanoPhotometer NP80 Specifications

General IconGeneral
BrandImplen
ModelNanoPhotometer NP80
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals