NanoPhotometer
®
N120/NP80/N60/N50/C40 User Manual Version 4.3.1
95
Note: Only available for cuvette applications (NP80 and C40).
For NanoVolume Application:
Select the dilution depending on the sample concentration.
Note: There is no automatic path length setting available in this method. Select either a
virtual dilution of 15 (path length 0.67 mm) or of 140 (path length 0.07 mm).
For NanoVolume application with submicroliter cell (C40 only):
Select Lid depending on your sample concentration (for concentration ranges of different lids
see page 52 Submicroliter Cell Measurements Basics).
Required sample volume for each lid is listed to the right.
Note: A new Blank is recommended when changing the dilution lid.
4. Default wavelength () is 340 nm but can be changed in the range of
200–900 nm (N50: 200–650 nm), depending on the application.
5. Time settings:
Enter the duration time in minutes over which measurements are to be taken. Possible
range is 1–3000 minutes.
Enter the interval time between measurements in seconds. Possible interval times are 5-
3,600 seconds (N50: 10–3,600 sec.), depending on the duration time.
Enter the delay time in seconds before the first measurement is taken. Possible delay time
is between 0–3,600 seconds, depending on the duration time.
Note: A maximum of 500 samples is possible. Please consider this when choosing the
duration and interval time.