Implen OD600
®
User Manual Version 1.0
16
When selecting a set, all samples within that set are automatically included.
Four samples/sets can be selected at once. Selected samples can be deleted or exported for further analysis. Use
the ALL button to quickly manage selections.
If your device enters deep sleep mode, a new set will be automatically started.
Note: If the data is selected manually, be aware that only the four samples displayed on the screen will be selected. The
selection will be lost after scrolling down to the next samples.
3. Exporting Results:
Please insert a USB storage device to proceed with the export.
Select sets/samples to export, then press the EXPORT button.
You can export your data in the following formats: .csv or .xls (default). Once a format is selected, tap the EXPORT
button to initiate the export process.
The data will be automatically saved in a new folder created on the USB stick.
Note: When exporting data from the sample list, the Excel format allows for exporting fewer than 200 samples at a time.
For larger datasets, the CSV format should be used, as it does not impose any limit on the number of samples that can be
exported in a single operation.
Note: If the USB stick is not working, please re-format it to FAT32 using a Windows PC.
Sample Handling Tips
The cuvette holder is compatible with glass or plastic test tubes/falcon tubes (10, 12, 16, 18 mm diameter - no
adapters are required) as well as all standard macro or semi-micro cuvettes (quartz, glass and plastic cuvettes with
an optical height of 8.5 mm).
It is also possible to use cuvettes with 5 mm, 2 mm or 1 mm path lengths, but there may be an adapter necessary.
Please ask your cuvette supplier for suitable adapter.
The minimum volume for accurate measurements depends on the cuvette/tube type used; it is necessary for the light
path to pass through the sample for accurate measurements.
Note: Do not overfill the cuvette for risk or spillage.
Note: The cuvette holder is not removable.
Note: Do not use aggressive solvents such as strong acids or bases or organic solvents at any time. If unsure please
contact support@implen.de for detailed information about your specific reagent/solvent.
Note: Highly concentrated acids and bases are not recommended. It is recommended to wipe the sample surface with a
lint-free laboratory wipe immediately upon completion of each measurement. For more information about compatibility of
specific solvents not listed above, please contact the Implen support team (support@implen.de) to check the compatibility.