Power Management Features
Battery Operation
A full battery pack can be operated for approximately 8 hours within normal usage. When the battery gets low there
will be a first warning message “Battery low”. It is recommended to charge the battery at this status. If the battery is
not charged there will be a second warning message when the battery is empty before the instrument will
automatically shut down.
The battery pack is designed for at least 500 full charging cycles. Afterwards the capacity and operation time of the
battery may vary. A fully charged battery pack will self-discharge (switched off device) over a time of 10 – 14 days.
After this time recharging of the battery pack is necessary.
Note: If the battery status has not reached 100% after the maximum charging time, keep the device turned on and connect
it to the main power. The battery status should then update.
Auto-Shutdown Features
If the device is not charged and remains idle on a desk without any operation, it will follow a specific power-saving
sequence:
After 10 minutes: The screen will automatically dim to conserve energy.
After 19.5 minutes: A popup notification will appear, stating that the device is about to shut down.
After 20 minutes: The device will shut down automatically if no operation has occurred.
However, if the device is connected to a charger while not in use, it will remain powered on indefinitely, without
initiating the shutdown sequence.
Note: If the Implen OD600
®