9. Maintenance
Maintenance-free technology
The Implen OD600
®
technology is maintenance-free. Regular maintenance and calibration is not necessary.
For facilities that are working according to national as well as international guidelines and standards including: Good
Laboratory Practice (GLP), Good Manufacturing Practice (GMP), or ISO9000-9004; the proper performance of the
device has to be tested and proven on a regular basis with individually set intervals. A certified Neutral Density filter
can be used as an optional accessory. These filters are suitable for the control and documentation of the photometric
accuracy of your system. Please contact your local Implen office or an authorized Implen partner for further
information.
Support agreements that help to fulfill the demands of regulatory guidelines concerning GLP/GMP include:
calibration certification using filters traceable to international standards (during production and quality control),
certified engineers and calibrated test equipment, approved to ISO 9001 standard, automatic self-diagnostic
calibration test during start of the Implen OD600
®
, result is documented in each data output file, and users are given
the possibility to save a Implen Document Source (IDS) file (no data manipulation possible).
Software update
There is a button for software update. It is needed to plug in a USB flash drive with the software installation file to the
USB port. The software update files are available on the Implen homepage: www.implen.de.
Update procedure:
1. Copy the Software Update file to a USB flash drive
2. Plug in the USB flash drive to the USB port of the Implen OD600
®
device
3. Go to Settings > Setup > Software Update
4. Press the Update button. The update will start and once finished automatically reboot the device.
Note: Make sure that the OD600 is connected to power and that the power connection is not interrupted during the update.
Note: Make sure that the sample name is NPOS.bin without any additions or extensions like from multiple downloads,
otherwise rename file to be NPOS.bin exactly.
Contact Implen
There is an option to contact Implen for instruments connected to the internet. For any support issues and questions
please contact the Implen team directly:
Europe, Asia, South Pacific, North and South America
Middle East, Africa
Phone: +49 89 72637180 Phone: +1 818 748 6400
Fax: +49 89 726371854 Fax: +1 818 449-0416
Email: support@implen.de Email: support@implen.com
Website: www.implen.de Website: www.implen.com
Implen GmbH Implen, Inc.
Schatzbogen 52 Unit 104
81829 München 31194 La Baya Drive
Germany Westlake Village, CA 91362
USA

Table of Contents

Related product manuals