Installation 33
iina70e 02.fm technical handbook(1408)
2.5 Default parameters
The following parameters are set like shown when in the menu of the UL1000 and
UL1000 Fab under Settings Parameters load/save, „load default values“ is
chosen.
Auto-scaling: On
Scaling logarithmic
Display range: 4 decades
Time axis: 32 seconds
LCD invers: OFF
Background in stand by mode: OFF
Automatic calibration request: OFF
Mass: 4 (helium)
Recorder Output: leak rate
Volume: 2
Leak rate unit: mbar l/s
Mode: Vacuum
Trigger level 1: 1E-9 mbar l/s
Trigger level 2: 1E-8 mbar l/s
Leak rate external test leak (Vacuum): 1E-7 mbar l/s
Leak rate external test leak (Sniffer): 1E-5 mbar l/s
Vent delay: 2 seconds
Automatic purge: OFF
Pressure unit: mbar
Minimum volume: 0
Beep: ON
Maximum evacuation time: 30 minutes
Audio Alarm Typ: Trigger Alarm
Maximum inlet pressure when sniffing: 1 mbar
Minimum Inlet pressure when sniffing: 0,1 mbar
Number of decimal place at leak rate displayed: 1
Scroll display: On
Particle protection: Off
Direct access to calibration: On
Contamination protection: Off
Switch off limit for contamination protection: 1E-3 mbar l/s
Control location: Local
Alarm delay: 30 seconds
Leak rate filter: I•Cal
Zero: enabled