Description of the Menu 65
iina70e 06.fm technical handbook(1408)
6.4.6.3 Setpoint
The tone is off as long as the leak rate is below the Trigger level 1. Above Trigger 1
the tone varies proportional to the leak rate Fig. 6-11.
6.4.6.4 Trigger alarm
As soon as the leak rate increases above trigger level 1, a multi-tone signal is
generated. The tone does not vary with the leak rate.
6.5 Calibration
Main Menu > Calibration
Please refer to Chapter 7 Calibration for a detailed description of the calibration.
Fig. 6-11: Setpoint
f
LRTr. 1
no tone
375 Hz -
1000 • TR.1
3000 Hz -
Fig. 6-12: The Calibration Menu