92 Calibration
iina70e 07.fm technical handbook(1408)
The bargraph display shows a signal which must not decrease any more. There
might be a small fluctuation which is okay. If so please press OK (Soft Key no. 8).
The UL1000 and UL1000 Fab displays the old and the calculated new calibration
factor.
Fig. 7-7: External Calibration, Step 7
Fig. 7-8: External Calibration, Step 8