- 43 -
STATUS/ERROR MESSAGES
Messages Descriptions
Sensor Off
The CO
2
sensor is not connected
Sensor Warm Up
One of the following conditions exist:
Sensor under temperature
Temperature not stable
Source Current unstable
Sensor Over Temp
Make sure sensor is not exposed to extreme heat (heat lamp,
etc.). If error persists, return sensor to factory for servicing.
Sensor error
Check that the sensor is properly plugged in. Reinsert or reset
the sensor if necessary. If error persists, return sensor to factory
for servicing.
Sensor Zeroing. . .
A zero is currently in progress.
Zero Required
To clear, check airway adapter and clean if necessary. If this
does not correct the error, perform an adapter zero. If you must
adapter zero more than once, a possible hardware error may
exist.
Check Sampling Line
To clear, clean if sampling line mucus or moisture is seen. If
the sampling line is clean, perform a zero.
CO
2
Out of Range
The value being calculated is greater than the upper CO
2
limit
(150 mmHg, 20.0 kPa, or 19.7 %). The maximum value output
is the upper CO
2.
Check Airway
Adapter
To clear, clean airway adapter if mucus or moisture is seen. If
the adapter is clean, perform a zero.
Pump Life Exceed
The manufacturer stated pump life has been exceeded. Service
may
be required if Pneumatic System Error is present and can
no longer be cleared.
Sensor Setup. . .
The CO
2
sensor is setting process.
EtCO
2
Zero Error:
Sensor Not Ready.
The CO
2
sensor is not ready for a EtCO
2
Zero
EtCO
2
Zero Error:
Breath Detected.
Breaths have been detected by the CO
2
module within the last
20 seconds while a CO
2
module zero was attempted.
MAINTENANCE AND CLEANING
SCHEDULE
The CO
2
Module flow rate accuracy should be verified by direct measurement using a
calibrated flow meter every 12 months.
CLEACING
Cleaning the CO
2
Module case, Cable and connector:
1. Use a cloth dampened with isopropyl alcohol 70%, a 10% aqueous solution of sodium
hypochlorite (bleach), a 2% gluteraldehyde solution, ammonia, mild soap or disinfectant
spray cleaner such as Steris Coverage® Spray HB.
2. Wipe down with a clean water-dampened cloth to rinse and dry before use. Make
certain that the sensor windows are clean and dry before reuse.
Cleaning the Sidestream On-Airway Adapters and Sidestream Sampling Kits:
Sidestream on-airway adapters and sidestream sampling kits are single patient use. Treat
in accordance with hospital protocols for handling single patient use devices.
[NOTE]: DO NOT immerse or sterilize the CO
2
Module.