User’s manual of CLEO Patient Monitor
- 48 -
BATTERY OPERATION
CLEO Patient Monitor is designed to operate on one rechargeable Lithium ion battery
whenever AC power supply is interrupted. A symbol is displayed in the upper right quarter
of the screen to indicate the status of recharging, in which the color part represents the
electric energy of the battery.
A new, fully charged battery will provide about 3 hour of monitoring time under the
following conditions: no audible alarms, no analog or serial output devices attached, and
no backlight. The charge and discharge cycles life of the battery is about 300 times.
When operating on battery, the monitor will shut off automatically when the electric energy
is low. When the electric energy is lower than 25% of total power capacity, the battery
signal will change red. Connect the monitor to AC power at this moment can recharge the
battery while operating. If keep operating on the battery, the monitor will shut off
automatically upon exhaustion of the battery.
BATTERY RECYCLE
When a battery has visual signs of damage, or no longer holds a charge, it should be
replaced. Remove the old battery from the Patient Monitor and recycle it properly. To
dispose of a battery, follow local laws for proper disposal.
[NOTE]: Whenever the monitor is connected to AC power, the battery is being
charged. Therefore, it is recommended that the monitor remain connected to AC
power when not in use. This will make available a fully charged battery for use at any
time.
[NOTE]: As the battery is used and recharged over a period of time, the amount of
time between the onset of the low battery alarm and the instrument shut-off may
become shorter.
If the backlight is turned off during a low battery condition, it cannot be turned back on.
It is recommended that the internal battery is replaced by qualified service personnel
every 24 months.
[CAUTION]: If the CLEO is to be stored for a period of 2 months or longer, notify
service personnel to remove the battery from the monitor prior to storage. Recharge
the battery when it has not been charged for 2 or more months.
[WARNING]: DO NOT disassemble batteries, or put them into fire, or cause them to
short circuit. They may ignite, explode, or leak, causing personal injury.