User’s manual for CLEO Patient Monitor
- 54 -
EMC
The product is in radio-interference protection class A in accordance with EN55011. The
product complies with the requirements of standard EN60601-1-2:2007 “Electromagnetic
Compatibility - Medical Electrical Equipment”.
ELECTROMAGNETIC IMMUNITY
This section constitutes the guidance and CLEO Patient Monitor’s declaration regarding
electromagnetic immunity. The CLEO Patient Monitor is intended for use in the
electromagnetic environment specified below. The customer or the user of the CLEO
Patient Monitor should ensure that it is used in such an environment.
Electromagnetic environment
– guidance
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV a ir
6 kV contact
8 kV a ir
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, th
e relative humidity
should be at least 30%.
transient/burst
IEC 61000-4-4
supply lines
1 kV for input /
supply lines
1 kV for input /
Mains power quality should be that of a
typical c
environment.
IEC 61000-4-5
differential
Mode
differential
Mode
differential
Mode
differential
Mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5 % UT
1
(>95 % dip in
UT) for 0.5
cycle
40 % UT
cycles
70 % UT
cycles
<5 % UT
(>95 % dip in
UT) for 5 sec
UT) for 0.5 cycle
40 % UT
for 5 cycles
70 % UT
for 25 cycles
<5 % UT
(>95 % dip in
UT) for 5 sec
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the CLEO
Patient Monitor requires continued
operation during power mains
interruptions, then the CLEO
Monitor should
uninterruptible power supply or a battery.
(50/60 Hz)
magnetic field
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels typical in a commerc
hospital environment.
Note:
UT is the a. c. mains voltage prior to application of the test level.