User’s manual for CLEO Patient Monitor
- 53 -
ETCO
2
Calculation:
Method: Peak of the expired CO
2
waveform
Selections: 1 breath, 10 s, 20 s
Inspired CO
2
Measurement:
Range: 3
50 mmHg
Method: Lowest reading of the CO
2
waveform
in the previous 20 s
Selection: 20 s (not user-selectable)
awRR Measurement Range:
2
150 rpm
awRR Accuracy:
±1 breath
EtCO
2
Alarm Limit:
Upper Limit : 20
100 mmHg
Lower Limit : 10
95 mmHg
awRR Alarm Limit:
Upper Limit : 10
150 rpm
Lower Limit : 5
100 rpm
Apnea Time:
10
60 s
Anesthetic Agent Effects
(MAC levels)
Anesthetic Agent
Sensitivity
1)
(uncompensated)
Accuracy specification will
be maintained for
halogenated anesthetic
agents present at accepted
MAC (Minimum Alveolar
Concentration) clinical
levels.
1)
MAC Level %(v/v): Halothane 0.74; Enflurane 1.68; Isoflurane 2.00; Desflurane 6.30.
NETWORKING
Wired Networking:
Industry Standard: IEEE 802.3 wired network
RJ45 interface and USB port
POWER
Source:
External AC Power and Internal Battery
AC Power:
100/240 VAC, 50/60 Hz, 30VA, F2AL 250V
Battery:
Rechargeable Lithium ion battery, 7.2 V/2000
mAh
Operating time under
normal conditions:
3 hrs
Operating time after
first
sounding of
low-battery alarm:
10 mins
Charge Time:
8 hrs
Number of Batteries:
1
ENVIRONMENTAL SPECIFICATIONS
Temperature:
Operating: 0
40 °C
Storage: -20
60°C
Humidity Range (Noncondensing):
Operating: 15
95%
Storage: 10
95%