4.5 Troubleshooting
Problem Probable cause Remedy
The connection
between pipette and
ASSIST PLUS can-
not be established.
The two instruments have not
been paired or the pairing
was lost.
From the main menu of the
pipette go to Toolbox -> ASSIST
PLUS and select Bluetooth
Pairing, see “3.9.3
Device
Information” on page 14. Follow
the instructions on the pipette
screen.
Error LED (18)
blinks red
ASSIST PLUS motor lost
steps during movement.
Pipetting heights were set
wrong and the pipette
crashed into the plate.
Follow the instructions on the
pipette display.
After firmware
update, the Error
LED (18
) blinks red
and ASSIST PLUS
cannot be started.
Firmware may be corrupted. Contact INTEGRA Bioscience
service.
The tip height is not
correct using 12.5,
300 or 1250 μl Grip-
Tips.
Wrong tip type set. Enter the tip type used, see
“3.9.5
Tip Monitor” on page 15.