ASSIST PLUS – Operating instructions V02 Programming
32 INTEGRA Biosciences AG
5.7.3 Multi aspirate program
Application: This program can be used for pooling applications or removal of
supernatants.
Options Steps Description of Multi aspirate
Plate Wells Sets the type of well plate (12, 24, 48, 96 or 384) to be used
as (source)/target. On VOYAGER pipettes this automatically
defines the tip spacing. Note: only pipette compatible plates
can be selected, see “7.6
Plate compatibility” on page 45.
Orientation Sets the orientation of the plate (landscape, portrait).
Note: orientation can only be changed if it is compatible with
the pipette.
Clear Height Sets the travel height at which the GripTips move above the
plates.
Single
Transfer
Instead of pipetting aliquots, a single transfer is performed.
Aspiration volume = Dispense volume.
Tip Change If activated, tips are changed before each new aspiration.
Source Plate Select the deck position (B or C) as location for the source
plate.
Count Sets the total number of aspiration steps.
First Column
or First Row
Define the column/row where the first volume should be
aspirated from.
Repeat
Volume
If activated, identical volumes are aspirated.
If off, define the different volumes for aspiration.
Aspirate
(1 to n)
Set the volume(s) to be aspirated (for every aspirate step).
Asp Speed Sets speed uniquely for aspiration (1 = low, 10 = fast).
Height Set the (start) Height for all aspiration (and mixing) steps.
Mixing Mixes the sample in the source container before each
aspiration.
If on, define Mix volume, Mix Speed and number of Mix
Cycles.