ASSIST PLUS – Operating instructions V02 Technical Data
www.integra-biosciences.com 47
7.8 Tip travel distances
The Tip Travel defines the distance which the tips move during an aspiration, dispense
and mix step to artificially track the liquid level. The Tip Travel starts from a user defined
height as shown below.
Section Steps and parameters
Source Source Mix (
):
Tip Travel starts from Height (H
s
).
Moves the Source Mix (S
m
)
distance down and up.
Source Asp. (
):
Tip Travel starts from Height/Start
Height (H
s
). Moves the Source Asp.
(S
a
) distance down.
Tar ge t Target Disp. (
):
Tip Travel starts from
Height (H
t
). Moves the
set Target Disp. (T
d
)
distance up.
Target Mix ():
Tip Travel starts after
dispense from the
Height (H
t
) plus Target
Disp. (T
d
). Moves the
Target Mix (T
m
)
distance down and up.
Target Asp. (
):
Tip Travel starts from
Height (H
t
) plus Target
Disp. (T
d
) distance.
Moves the Target
Disp. (T
d
) distance
down.
(H