ASSIST PLUS – Operating instructions V02 Programming
www.integra-biosciences.com 31
Press to save your settings. This will return you to the list of Repeat/Variable Dispense
programs.
Target TipTouch It is highly recommended to activate a tip touch after a
dispensing step to remove drops that may cling to the pipette
tips.
Sets where the tip touch should be performed.
“Liquid”: the tips will dip into the center of the wells.
“Side”: the tips will touch to the side of the wells.
Defines the height for the tip touch (“TipTouch at”).
Plate count Sets the total number of target plates (1-9) used in the
current run.
Post-
Dispense
A Post-Dispense volume can be selected independently,
which is discarded, to improve accuracy and precision.
Note: only available if Single Transfer is inactive.
If on, define:
the location where the Post-Dispense should be discarded
(ejected with tips or dispensed into the source container)
and
the Post-Dispense volume. Suggested: at least 3-5% of the
pipette's maximum volume.
Reuse
Post-Disp.
Only Repeat dispense:
If on, at the end of the program the post-dispense remains in
the tip, while the pipette is ready to aspirate a new volume to
start the next repeat dispense run.
Options Steps Description of Repeat/Variable Dispense