ASSIST PLUS – Operating instructions V02 Programming
30 INTEGRA Biosciences AG
Source
(Gene-
ral)
Start Height The height where aspiration starts. The ASSIST PLUS will
automatically refill the GripTips if the total dispense volume
required for the target exceeds the maximum volume of the
pipette. The first aspiration will be at the start height and for
consecutive aspiration steps the pipette will go lower until it
reaches the end height.
Note: only visible if Tracking is switched off.
End Height Sets the height for the last aspiration step.
Asp Speed Sets speed uniquely for aspiration (1 = low, 10 = fast).
Mixing Mixes the sample in the source container before each
aspiration.
If on, define Mix volume, Mix Speed and number of Mix
Cycles.
Target Plate Sets the destination plate position (B or C).
Pre-Dispense A Pre-Dispense volume can be selected independently,
which is discarded back to the source immediately after
aspiration, to improve accuracy and precision.
Note: option only available if Single Transfer is inactive.
If on, define the Pre-Dispense volume. Suggested: at least
3-5% of the pipette's maximum volume.
Count Sets the total number of dispensing steps.
First Column
or First Row
Sets the column/row where dispensing of the first dispense
should be started (column 1-24 or row A-P).
Dispense Only Repeat dispense:
Sets the volume to be dispensed in each well. The total
aspiration volume is calculated automatically. The pipette
cannot be overfilled.
Dispense
1 to n
Only Variable dispense:
Set the different dispense volumes for every variable
dispense step.
Disp. Speed Sets speed of all dispensing steps (1 = low, 10 = fast).
Height Sets the height for the dispensing steps.
Options Steps Description of Repeat/Variable Dispense