SERVICE INSTRUCTIONS
99
7.19.2 Repairing a flat tyre (type 3.10/4.50-6 pneumatic tyres)
Pre-requisites:
Flat screwdriver
10 mm Allen key
24 mm open-ended spanner
Repair kit for tyres with inner tube or a new hose.
Talcum powder
Removing the wheel
Jack up the vehicle (Place a piece of wood under
the frame).
Carefully remove the cover cap (1) using the flat
screwdriver.
Use the Allen key to secure the axle (2) and
loosen the nut on the inner side of the rocker
(covered) with the open-ended spanner.
Pull off the wheel and the axle in an outward
direction.
PLEASE NOTE
Assembly is carried out in the reverse order. Please ensure that the wheel is re-fitted on the
same side and in the same running direction it was in before removal.