EasyManua.ls Logo

Invacare Robin - Service Manual

Invacare Robin
228 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Invacare®Robin™
Robin™,RobinMover™
enCeilingHoist
ServiceManual..............................3
daLøftemotor
Servicemanual...............................33
frLève-personnesurrail
Manueldemaintenance......................65
itSollevatoreasoffitto
Manualeperlamanutenzione.................99
nlPlafondlift
Servicehandleiding...........................133
no
Takmontertløfter
Servicehåndbok.............................165
svTaklyft
Servicemanual...............................197
DEALER:Keepthismanual.
TheproceduresinthismanualMUSTbeperformedbyaqualifiedtechnician.

Table of Contents

Other manuals for Invacare Robin

Questions and Answers

  • S
    Shannon SmithSep 12, 2025
    What to do if Invacare Personal Care Products ceiling hoist does not respond to the hand control’s buttons and up/down buttons on the ceiling hoist are not responding?
    • J
      Jessica StewartSep 12, 2025
      If the Invacare ceiling hoist isn't responding to the hand control or the up/down buttons, it could be due to several reasons. First, check if the emergency stop is activated and reset it. Another possibility is that the system has no power or the battery is discharged; in this case, charge the battery or use the emergency battery. Lastly, a blown fuse could be the culprit, so replace the fuse.
  • A
    Alex AyersSep 12, 2025
    Why Invacare Medical Equipment ceiling hoist speed is the same with and without load?
    • B
      Brian ThompsonSep 12, 2025
      If the speed of your Invacare Medical Equipment ceiling hoist is the same with and without a load, contact an Invacare representative or specialized dealer for service. The non-load quick movement function may be permanently activated. Alternatively, clean or replace the hooks, as the lifting hooks may not turn freely enough; for this, contact an Invacare representative or specialized dealer for service.
  • E
    Emily BowmanSep 10, 2025
    Why Invacare Medical Equipment ceiling hoist does not traverse the rails?
    • J
      jwilsonSep 10, 2025
      If your Invacare Medical Equipment ceiling hoist does not traverse the rails, contact an Invacare representative or specialized dealer for service. The rails may require service or cleaning. Also, if you have a Robin™ model, the trolley might be worn or damaged and need replacement. If you have a Robin Mover™ model, the driving wheels may be worn or damaged and need adjustment, or the hand control might be defective and need replacement.
  • E
    Eric SmithSep 9, 2025
    What to do if no indicator light on Invacare Medical Equipment hoist?
    • D
      dhawkinsSep 9, 2025
      If there is no indicator light on your Invacare Medical Equipment hoist, check that the ceiling hoist has a power supply and that the battery is recharged. Also, place the hand control in the charging station and recharge the ceiling hoist.
  • N
    natalie96Sep 8, 2025
    Why my Invacare Medical Equipment ceiling hoist makes excessive noise?
    • W
      wtorresSep 8, 2025
      If the Invacare Medical Equipment ceiling hoist emits excessive noise when activated, the hoist must be replaced. Contact an Invacare representative or specialized dealer for service.
  • J
    Jessica RichardsonAug 23, 2025
    What to do if Invacare Lifting Systems hoist is not responding?
    • R
      Raymond JoyceAug 23, 2025
      If your Invacare Lifting Systems hoist isn't responding to the hand control or the up/down buttons on the hoist, first check if the emergency stop is activated and, if so, depress the emergency stop button. Next, verify that the battery isn't discharged; if it is, put the hoist on charge or use the emergency battery. Finally, inspect the fuse and replace it if it's blown.
  • J
    John GuerraSep 6, 2025
    What to do if Invacare Medical Equipment lifting straps only move down?
    • A
      Alexis GreenSep 6, 2025
      If the lifting straps on your Invacare Medical Equipment will only move down and not up, it could be because the battery is nearly discharged. Try charging the battery or using an emergency battery. Also, the load might be too high. Reduce the load, as the hoist is equipped with an over-current safety system.
  • C
    coxdavidSep 5, 2025
    Why Invacare Medical Equipment ceiling hoist does not respond to any buttons?
    • Z
      zharrisSep 5, 2025
      If the Invacare Medical Equipment ceiling hoist does not respond to the hand control’s buttons and the up/down buttons on the ceiling hoist are also not responding, it could be due to several reasons. First, check if the emergency stop is activated and reset it if necessary. Second, ensure that the system has power or the battery is adequately charged. If not, charge the battery or use an emergency battery. Lastly, the fuse might be blown, requiring replacement by contacting an Invacare representative or specialized dealer.
  • C
    colebobbySep 4, 2025
    What to do if Invacare Medical Equipment ceiling hoist responds to up/down buttons but not hand control?
    • M
      mandygarzaSep 4, 2025
      If the Invacare Medical Equipment ceiling hoist does not respond to the hand control buttons, but the up/down buttons on the ceiling hoist are working, the hand control might not be connected properly. Check the hand control plug to ensure it is properly inserted. Alternatively, the hand control itself may be defective and require replacement.
  • M
    Michael JohnsonSep 3, 2025
    How to fix Invacare Medical Equipment ceiling hoist that will not charge?
    • J
      Jeanette MurphySep 3, 2025
      To address this issue with your Invacare Medical Equipment ceiling hoist, first check that the power supply is switched on and connected to the charging station. If this does not solve the problem, contact an Invacare representative or specialized dealer for service.

Summary

General Information

Safety Information

Highlights warnings and cautions for safe operation and maintenance of the hoist.

Service

Installing the Ceiling Hoist to the Rail System

Details the process of installing the Invacare Robin hoist onto the rail system.

Replacing the Hand Control

Detailed steps for replacing the hand control unit of the hoist.

Replacing Strap and Hook Assemblies

Guides on replacing the lifting straps and hook components of the hoist.

Replacing the Battery

Step-by-step instructions for replacing the hoist's battery pack.

Replacing the PCB (Printed Circuit Board)

Detailed procedure for replacing the Printed Circuit Board (PCB) inside the hoist.

Replacing the Fuse

Instructions for safely replacing the fuse in the hoist unit.

Adjusting Loose Strap Switch

Guide on how to adjust the loose strap switch for proper function.

Replacing the Twin Speed Switch

Steps for replacing the twin speed switch component within the hoist.

Replace Contact Disc for Robin Mover

Instructions for replacing the contact disc on Robin Mover hoist models.

Maintenance Procedures

Safety Inspection Checklist

Comprehensive checklist for regular safety inspections of the hoist and rail system.

Troubleshooting Guide

Troubleshooting Table

A table listing common hoist problems and their corresponding solutions.