Configuration Commands
Copyright © Itech Electronic Co., Ltd. 19
Chapter3 Configuration Commands
CONFig[:SOURce]:VOLTage[:LEVel]:MINimum
This command is used to configure the voltage lower limit, same as the Volt-Min
setting in the instrument panel menu.
Command syntax
CONFig[:SOURce]:VOLTage[:LEVel]:MINimum
Parameters
MINimum|MAXimum|Float
Example
CONF:VOLT:MIN 2
Query syntax
CONFig[:SOURce]:VOLTage[:LEVel]:MINimum?
CONFig[:SOURce]:VOLTage[:LEVel]:MAXimum
This command is used to configure the voltage upper limit, same as Volt-Max
setting in the instrument panel menu.
Command syntax
CONFig[:SOURce]:VOLTage[:LEVel]:MAXimum
Parameters
MINimum|MAXimum|Float
Example
CONF:VOLT:MAX 2
Query syntax
CONFig[:SOURce]:VOLTage[:LEVel]:MAXimum?
CONFig[:SOURce]:FREQuency:MINimum
This command is used to configure the frequency lower limit, same as Freq-Min
of instrument panel menu.
Command syntax
CONFig[:SOURce]:FREQuency:MINimum