List Operation Commands
Copyright © Itech Electronic Co., Ltd. 40
LIST:STEP:FREQuency
This command is used to sets the LIST single-step frequency.
Command syntax
LIST:STEP:FREQuency
Parameter 1
Number of steps, 0 ~ 99
Parameter 2
Frequency value, Float
Example
LIST:STEP:FREQ 1,60
Query syntax
LIST:STEP:FREQuency?
Parameters
Number of steps, 0 ~ 99
Example
LIST:STEP:FREQ? 1
LIST:STEP:SLOPe
This command is used to set the LIST single step slope time. Parameters 1 and
2 are mandatory.
Command syntax
LIST:STEP:SLOPe
Parameter 1
Number of steps, 0 ~ 99
Parameter 2
Slope time, Float, Unit: mS
Example
LIST:STEP:SLOP 3,5
Query syntax
LIST:STEP:SLOPe?