TRACe Subsystem Commands
Copyright © Itech Electronic Co., Ltd. 30
Command syntax
TRACe:POINts <NRf+>
Parameters
2 to 1000 | MINimum | MAXimum | DEFault
Example
TRAC:POIN 10
Query syntax
TRACe:POINts? [MINimum|MAXimum|DEFault]
Returns
<NR1>
Related commands
TRAC:FEED
TRACe:FEED
This command is used to select the source of readings to be placed in the
buffer. With VOLTage selected, voltage readings are placed in the buffer. With
CURRent selected, current readings are placed in the buffer. With TWO
selected, voltage and current are placed in the buffer when storage is
performed, TRAC:POIN maximum values is 1000.
Command syntax
TRACe:FEED <CRD>
Parameters
VOLTage|CURRent|TWO
Example
TRAC:FEED VOLT
Query syntax
TRACe:FEED?
Returns
<CRD>
Related commands
TRAC:POIN