Calibration Commands
Copyright © Itech Electronic Co., Ltd. 54
Parameters
P1|P2
CALibrate:POWer [:DATa] {<numeric value>}
This command returns the actual output power value of the current calibration
point of the power supply.
Command syntax
CALibrate:POWer [:DATa] <NRf>
Parameters
<NRf>
Example
CAL:POW 80W
CALibrate:STRing
This command is used to set the calibration information during calibration.
Command syntax
CALibrate:STRing <Parameters>
Parameters
Support inputting a maximum length of 24 letters of string, which is recorded
the user calibration information. Such as the calibration time, frequency and so
on.
Example
CAL:STR "2005-1-9 20:12"
CALibrate:STRing?
This command is used to check the current calibration information.
Query syntax
CALibrate:STRing?
Returns
Calibration information saved in the power supply.