Technical Specifications
Copyright © ITECH Electronic Co., Ltd. 62
Model IT8930-150-1920
Rated value
(0
40
)
operating
0.31V/192A 3.1V/1920A
Constant
voltage
mode
Range 0.1~18V 0.1~150V
Resolution 1mV 10mV
Accuracy ±(0.05%+0.05%FS) ±(0.05%+0.05%FS)
Constant
current
mode
Range 0~192A 0~1920A
Resolution 10mA 100mA
Accuracy ±(0.1%+0.1%FS) ±(0.2%+0.1%FS)
Constant
resistance
mode *1
Accuracy
0.01%+0.08S
0.01%+0.0008S
Constant
power
Mode *3
Measuring range
Read-back
voltage
Range 0~18V 0~150V
Resolution 1mV 10mV
Accuracy ±(0.025%+0.025%FS) ±(0.025%+0.025%FS)
Read-back
current
Range 0~192A 0~1920A
Resolution 10mA 100mA
Accuracy ±(0.1%+0.1%FS) ±(0.2%+0.1%FS)
Read-back
power
Range 30KW
Resolution 1W
Accuracy ±(0.2%+0.3%FS)
≒30.3KW
≒210A ≒2100A
≒157V
Overtemperat
ure protection
≒85℃
Short circuit
Current (CC)
≒200A ≒2000A
Voltage (CV) 0V 0V
Resistance (CR)
≒1.6mΩ ≒1.6mΩ
≒300KΩ
2
U
*1.The voltage/current input is no less than 10%FS (FS indicates the full
scope).
*2.The scope of read-back resistance is
(1/(1/R+(1/R)*0.01%+0.08),1/(1/R-(1/R)*0.01%-0.08)).
*3.The voltage/current input is no less than 10%FS.