EasyManua.ls Logo

Kamstrup MULTICAL 401 - Calibration and verification

Kamstrup MULTICAL 401
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TECHNICAL DESCRIPTION MULTICAL
®
401
5512-091 GB/02.2008/Rev. J1
55
13 Calibration and verification
In order to be able to carry out test/verification of MULTICAL
®
401 with a minimum use of time the meter has a
testing mode. When the meter is in testing mode the program runs approx. 4 times faster than in standard mode.
Testing mode also has some extra functions as described below.
(NB. MULTICAL
®
401 uses approx. 4 times more current in testing mode. However, under normal circumstances
the meter will only be in testing mode e.g. 9 hours per 5 years, which has no influence on the total lifetime of the
battery)
Figure 46: Module and test plug (No. 1-8 is sealed by a verification seal)
Flow verification
Flow measurement operates with absolute time measurement, and is therefore independent of the temperature
sensors. The flow part can be verified as a separate unit, if a 3-part verification is required.
STARTING UP
16 seconds must ellapse from start up to calibration in order to allow a true reading to be reached.
MEASURING FLOW
To obtain a correct flow measurement, calibration must take min. 2 minutes.
EVACUATION
MULTICAL
®
401 must NOT be evacuated (subjected to vacuum).
Calculator verification
The calculator can be verfied separately by means of temperature simulation resistances connected to the
terminal strips Nos. 5-6-7-8. Volume simulation can be made by means of the built-in ”Autointegration” (see
section 13.2).
Temperature simulation via the test plug is only possible when verification equipment type 66-88-385 is used
(see section 14.3
Verification with METERTOOL).
GND
tF temperature
tF sense
GND
GND
Tx1 – from
P to module
Input B or CE – Tx2
3.6 V internal supply
tR temperature
tR sense
Test input for
P
3.6 V supply for modules
Rx1 – from module to
P
Input A – Rx2

Table of Contents

Related product manuals