This limitation means that the legal parameters program, type and serial Nos. cannot be changed, while all other
data can be programmed as required.
This limitation is used to prevent the original operating parameters from being changed on type approved and
verified meters.
National verification demands must be checked before the calculator’s verification seal is broken.
14.2.2
Complete programming
If the programming lock is short-cirnuited, it is possible to reprogram MULTICAL
®
401, including the legal data
program, type and serial Nos. When data cable 66-99-108 or verification equipment 66-99-385 is used, the
programming lock is short-circuited.
Please note that the verification seals are automatically broken, when the test plug is used.
Sealing must be carried out by an autorised laboratory.
Please note that the data logging memory in the calculator cannot be changed/erased during programming,
unless this is selected in the software.
14.2.3
File
Under the menu “File” one of the following functions can be selected:
Open Customer Retrieves stored customer settings from the database
Save Customer Saves new customer settings in the database
Print Certificate Starts printing test certificate
Print Label Starts printing front label
Print Setup Printer setup for printing front label and certificate
Exit Terminates METERTOOL
14.2.4
Utility
This menu gives access to the following dialog boxes:
Programming General view which is used when reading and programming
Time/Date The PC’s date and time are transmitted to MULTICAL
®
401
Telephone No. 2 different telephone numbers can be programmed in MULTICAL
®
401
Preset VA/VB Presetting of input A/B values. The position of the decimal point depends on the
meter's FF/GG coding.
M-Bus address The primary M-bus address of the meter can be entered here independent of the
meter's customer number.
Info Code Is used for reading the info code as well as date, energy and volume at the time
when the info code occurred
Meter type Reads the meter’s internal software revision
Reset Resets all registers if the programming lock is short-circuited
Verification See section 14.3
Verification via METERTOOL
Flow meter Adjustment Is only used when adjusting the MULTICAL
401 flow part (requires a
password)
Password can be disclosed by Kamstrup A/S