ERGOcom 3
8 Troubleshooting
8 Troubleshooting
Fault Cause Remedy
No status display.
ERGOcom 3 switched to standby.
▶ Leave Standby mode.
See also: 5.1.2 Standby, Page 24
ERGOcom 3 off.
▶
ERGOcom 3 on.
Mains fuse burnt out. ▶ Test mains fuse and replace
where necessary.
ERGOcom 3 switches itself off and
cannot be switched back on imme-
diately.
Overheating. ▶ Ensure good ventilation for the
ERGOcom 3 .
See also: 4 Start-up, Page 20
Everything has stopped working. Mains failure / overvoltage. ▶
ERGOcom 3 to be switched off
and then switched on again after
5 seconds.
Access to USB hardware does not
work under Windows XP.
The Microsoft patch needed to ope-
rate the USB port is not installed on
the target computer.
▶ Install the required patch.
This can be done while installing
the CCCDIS.
See also: 4.1.3 Connect PC , Page
22
Selected image source (e.g. DVD
player) is not displayed.
The DVD player is switched off, or
no DVD has been inserted.
▶ Check image source.
No camera image is displayed. Treatment unit switched off. ▶ Switch on treatment unit.
Poor video quality. Poor image material. ▶ Only use image material that is of
good quality.
Poor PC image source quality. Display setting for graphics card not
the same as the display format set-
ting on the ERGOcom 3 .
▶ Adjust display format and gra-
phics card setting.
See also: 5.1.5 Menu points,
Page 25
Poor image quality. Incorrect settings in the OSD menu
display interface.
▶ Improve the image quality using
the options in the OSD menu.
See also: 5.4 Operation, OSD me-
nu display interface, Page 46
55/56