Table 
2-5. 
dB 
Specifications for 
AC 
Volts 
(600n 
Ref) 
dB 
M 
Range 
200mV 
ZV-75OV 
de (Ref: 
60001 
InDut 
1mV 
to 
2mV 
(-58 
to 
-52dBm) 
2mV 
to 
20mV 
(-52 
to 
-32dBm) 
20mV 
to 
200mV 
(-32 
to 
-12dBm) 
200mV 
to 
750V 
1-12 
to 
+59.8dBm) 
dBm 
Measurements with 
600a 
Reference Impedance 
dBm is defined as decibels above 
or 
below a 1mW reference. 
The standard reference impedance of  the Model 197 is 6000. 
What  that  means is that  the Model 197 is designed to read 
OdBm  when the voltage needed to dissipate 1mW through a 
600a 
impedance is applied to the Model 197. That calculated 
voltage  level  is  0.7746V as derived  from  the  basic  power 
equation. 
~~ 
E=h’.R 
- 
E 
=q0--3 
W.6000 
E 
= 
0.77456V 
Thus with  a  6000 reference impedance the  Model  197 will 
read OdBm  whenever 0.7746v is applied. 
NOTE 
Do 
not confuse reference impedance with input 
impedance. The input  impedance 
of 
the instru- 
ment is not modified in the dB mode. 
To 
make dBm measurements referenced 
to 
6000, proceed as 
follows: 
1. 
Connect the test leads to the INPUT 
HI 
and 
LO 
terminals 
2. Select the ACV 
or 
DCV function. 
3. 
Select autorange for optimum resolution. 
4. 
Press the dB button. 
5. 
Connect the test leads to the voltage source. 
6. Make the dBm reading from the display. 
of  the Model 197. 
20Hz- 
10kHz 
2.00 
0.85 
0.18 
0.18 
Accuraci 
10kHz- 
2OkHz 
3.00 
1.10 
0.18 
0.18 
( 
k 
dBw 
50kHz 
ZOkHz- 
- 
2.00 
0.28 
0.28 
5OkHz- 
100kHz 
- 
- 
0.65 
0.50 
dBm Measurements with Other Reference Impedances 
dBm  measurements  can  be  made  with  other  reference im- 
pedances.  The  most  convenient  method  for  using  other 
reference  impedances  is  to  algebraically  subtract  the 
calculated dB offset for the desired reference impedance from 
the reading  on the display  of  the Model  197. Table 2-6  lists 
common reference impedances and the corresponding offset 
values. The following equation  can be used  to calculate the 
offset the offset for impedances not listed in  Table 2-6. 
New ref 
Z 
Offset (for dBm) 
= 
10 log---- 
600Q 
To 
make  dBm  measurements  referenced  to  another  im- 
pedance, proceed as follows: 
1. 
Choose the desired reference impedance. 
2. Calculate 
or 
look up the offset value in Table 2-6 for the 
3. 
Determine  dBm  at  the  desired  reference  impedance  as 
dBm (at ref 
Z) 
= 
197 reading 
- 
offset 
desired reference impedance. 
follows: 
Example: Make dBm  measurements references to  a 1000 
reference impedance. 
A. 
1000 not  listed  in  Table  2-6 
so 
the  offset  must  be 
calculated as follows: 
Offset 
= 
10 log 
(z 
) 
Offset 
= 
-7.78dB 
B. 
Subtract -7.78  from all subsequent displayed readings 
on the Model  197. 
2-1 1 
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com