2-7.
Comparison of Average and TRMS Meter Readings
Waveform
Half-Wave Rectified Sine
Full-Wave Rectified Sine
+
lo-
-
0
-m-
Square
+lo-:%
Rectified Square Wave
+
10-
-
00
Rectangular Pulse
+1°53L
O-N
k-trlT
ll:
'DUTY
CYCLE'
Triangular Sawtooth
Ac Coupled
Peak
Value
1ov
1
ov
1
ov
1
ov
1
ov
1
ov
1
ov
RMS
Value
7.07V
5.00V
7.07V
1o.oov
5.00V
1ov
5.77v
Average
Responding
Meter Reading
7.07V
3.53v
7.07V
11.1ov
5.55v
11.1V.
q
5.55v
Ac Coupled
TRMS
Meter
Reading
7.07V
5.00V
7.07V
1o.oov
5.00V
1ov
5.77v
Averaging
Meter
Percent Error
0%
29.4%
0%
11%
1 1
7c
1.11
-
1)
x
100%
3.8%
2-16