2100-900-01 Rev. D / September 2011 Return to Section Topics B-7
Model 2100 6 1/2-Digit Resolution Digital Multimeter User’s Manual Appendix B: Remote Interface Reference
MEASure:FREQuency? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a frequency measurement with the specified range and resolution. The reading
is sent to the output buffer. For frequency measurements, the meter uses only one "range" for all
inputs between 3Hz and 300kHz. With no input signal applied, frequency measurements return
"0.”
MEASure:PERiod? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a period measurement with the specified range and resolution. The reading is
sent to the output buffer. For period measurements, the meter uses only one "range" for all inputs
between 0.33 seconds and 3.3
µsec. With no input signal applied, period measurements return
"0."
MEASure:TEMPerature?
Preset and make and make a temperature measurement with an RTD. The reading is sent to the
output buffer.
MEASure:CONTinuity?
Preset and make a continuity measurement. The reading is sent to the output buffer. The range
and resolution are fixed at 1k and 5 1/2 digits respectively.
MEASure:DIODe?
Preset and make a diode measurement. The reading is sent to the output buffer. The range and
resolution are fixed at 1VDC with 1mA current source and 5 1/2 digits respectively.
The CONFigure command
CONFigure:VOLTage:DC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for DC voltage measurements with the specified range and
resolution. This command does not initiate the measurement.
CONFigure:VOLTage:DC:RATio {<range>|MIN|MAX|DEF },{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for DC:DC ratio measurements with the specified range and
resolution. This command does not initiate the measurement. The specified range applies to the
source signal, and autorange is selected for the reference signal.