2100-900-01 Rev. D / September 2011 Return to Section Topics B-17
Model 2100 6 1/2-Digit Resolution Digital Multimeter User’s Manual Appendix B: Remote Interface Reference
CALCulate:DBM:REFerence {<value>|MINimum|MAXimum}
Set the dBm reference value. Choose from: 50 ~ 8000 ohms.
CALCulate:DBM:REFerence? [MINimum|MAXimium]
Query the dBm reference value.
DATA:FEED RDG_STORE,{"CALCulate"|""}
Selects whether readings taken using the INITiate command are stored in the multimeter's internal
memory (default) or not stored at all. In the default state (DATA:FEED RDG_STORE,"CALC"), up
to 2000 readings are stored in memory when INITiate is executed. The MEASure? And
CONFigure commands automatically select "CALC.” With memory disabled (DATA:FEED
RDG_STORE,""), readings taken using INITiate are not stored. This may be useful with the Min/
Max operation since it allows you to determine an average of the readings without storing the
individual values. An error will be generated if you attempt to transfer readings to the output buffer
using the FETCh? command.
DATA:FEED?
Query the reading memory state. Returns "CALC" or "".
The triggering commands
INITiate
Change the state of the triggering system from the "idle" state to "wait-for-trigger" state. The meter
will start to make measurements when a required triggering condition is met after the INITiate
command is received. Readings are stored in memory until you are ready to read them. Use the
FETCh? command to retrieve your measurements results.
READ?
Change the state of the triggering system from the "idle" state to the "wait-for-trigger" state. The
meter will start to make measurements when a required triggering condition is met after the
READ? command is received. Readings are sent to the output buffer immediately.
TRIGger:SOURce {BUS|IMMediate|EXTernal}
Select a trigger source. The multimeter will accept a software (BUS) trigger, an immediate internal
trigger, or a hardware trigger from the rear-panel EXT TRIG terminal.
TRIGger:SOURce?
Query the trigger source.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2100 and is the answer not in the manual?

Keithley 2100 Specifications

General IconGeneral
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ContinuityYes
Diode TestYes
Frequency Range3 Hz to 300 kHz
DC Voltage Accuracy0.0035%
DC Current Range100 µA to 10 A
DC Current Accuracy0.05%
AC Current Range100 µA to 10 A
Frequency Accuracy0.01%
Capacitance Range1 nF to 10 mF
Temperature Range-200 °C
ConnectivityRS-232
Type6.5-digit digital multimeter

Related product manuals