4-44 Return to Section Topics 2100-900-01 Rev. D / September 2011
Section 4: Front Panel Operations Model 2100 6 1/2-Digit Resolution Digital Multimeter User’s Manual
624 Unable to sense line frequency: The supplied voltage AC2 is routed through a comparator
U4 to generate a logic input signal. This test checks that the logic input from U4 to panel MCU U3
is toggling. If no logic input is detected, the instrument will assume 50Hz line operation for all future
measurements.
650 Front-End MCUs Firmware Version Error (System Error)
660 Firmware Version Error
661 External Memory Verion Error
662 Calibration Version Error
663 Scanner card Version Error
664 Temperature Version Error
665 GPIB Version Error
666 RS-232 Version Error
668 Manual Calibration Version Error
669 Self-Test Version Error
670 Command Version Error

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2100 and is the answer not in the manual?

Keithley 2100 Specifications

General IconGeneral
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ContinuityYes
Diode TestYes
Frequency Range3 Hz to 300 kHz
DC Voltage Accuracy0.0035%
DC Current Range100 µA to 10 A
DC Current Accuracy0.05%
AC Current Range100 µA to 10 A
Frequency Accuracy0.01%
Capacitance Range1 nF to 10 mF
Temperature Range-200 °C
ConnectivityRS-232
Type6.5-digit digital multimeter