SCPI Tables 12-17
:SOURce2
:CURRent
[:LIMit] Path to configure current limit.
[:VALue] <NRf> Specify current limit value in amps: 0.006–5
with 10
0μA resolution.
0.25
[:VALue]? Query current limit.
:TYPE <name> Select current limit type: LIMit or TRIP. LIM
:TYPE? Query current limit type.
:STATe? Query state of current limit: 1 = in current
l
imit (for LIMit type) or output tripped (for
TRIP type), 0 = not in LIMit/TRIP.
Table 12-6
STATus command summary (refer to Section 8)
Command Description
Default
parameter
SCPI
:STATus (Note 1)
:MEASurement Path to control the measurement event registers:
[:EVENt]? Read the event register. (Note 2)
:ENABle <NRf> Program the enable register. (Note 3)
:ENABle? Read the enable register.
:CONDition? Read the condition register.
:OPERation Path to control the operation status registers:
[:EVENt]? Read the event register. (Note 2)
:ENABle <NRf> Program the enable register. (Note 3)
:ENABle? Read the enable register.
Notes:1.Commands in this subsystem are not affected by *RST. The effects of cycling power, *CLS, and :STATus:PRESet
are explained by the following notes.
2. Event registersPower-up and *CLS — Clears all bits of the registers
:STATus:PRESet — No effect
3. Enable registersPower-up and :STATus:PRESet — Clears all bits of the registers
*CLS — No effect
4. Error queue Power-up and *CLS — Clears the error queue
:STATus:PRESet — No effect
5. Enable/disablePower-up — Clears list of messages
error queue *CLS and :STATus:PRESet — No effect
messages
Table 12-5 (cont.)
SOURce command summary (cont.) (refer to Table 2-3)
Command Description
Default
parameter
SCPI