14-16 SCPI Reference Tables Model 6487 Reference Manual
:COUNt? Read the number of messages in queue.
:CODE Code numbers only:
[:NEXT]? Read and clear oldest error/status (code
only).
:ALL? Read and clear all errors/status (codes only).
:SYSTem:
:CLEar Clear messages from error queue. 10
:KEY <NRf> Simulate key-press.
:KEY? Query the last “pressed” key.
RS-232 interface: 9
:LOCal While in LLO, removes the LLO and places the
Model 6487 in local (RS-232 only).
:REMote Places the Model 6487 in remote if not in LLO
(RS-232 only).
:RWLock Places the Model 6487 in local lockout (RS-232
only).
:KLOCK <b> When true, functions the same as local lockout
and prevents LOCAL key from taking unit out
of remote.
OFF 9
:KLOCK? Query key lock state.
Table 14-7 (cont.)
SYSTem command summary (see Section 13 for detailed information)
Command Description
Default
parameter
Ref
Section
SCPI

Table of Contents

Other manuals for Keithley 6487

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6487 and is the answer not in the manual?

Keithley 6487 Specifications

General IconGeneral
TypePicoammeter/Voltage Source
Current Measurement Range20 fA to 20 mA
InterfacesGPIB, RS-232
Voltage Burden< 200 µV
Input Impedance>10^14 Ω
Power100 V / 120 V / 220 V / 240 V
Basic Current Measurement Accuracy0.4%