Model 6487 Reference Manual Calibration 16-7
Calibration procedure
The calibration procedure should be performed in the following order:
Preparing for calibration
Offset voltage calibration
Current calibration
Voltage source calibration
Entering calibration dates and saving calibration
Locking out calibration
Preparing for calibration
1. Turn on the Model 6487 and the calibrator; allow them to warm up for at least one
hour before performing calibration.
2. Press MENU, select CAL, then press ENTER. The instrument will display the
following:
CAL: VOFFSET
3. Use the up or down RANGE key to display the following:
CAL: UNLOCK
4. Press ENTER. The instrument will prompt for the calibration code:
CODE? 000000
5. Enter the present calibration code on the display. (Factory default: 006487.) Use
the up and down RANGE keys to select the letter or number and use the left and
right arrow keys to choose the position. Press ENTER to complete the process and
the unit will display:
CAL UNLOCKED
Followed by:
NEW CODE? N
6. With N displayed, press ENTER.
Offset voltage calibration
Before performing the remaining calibration steps, perform offset voltage calibration as
outlined below.
1. From the calibration menu, use the up or down RANGE key to display the
following:
CAL: VOFFSET
2. Press ENTER. The instrument will prompt as follows:
INPUT CAP
3. Connect the triax shielding cap to the rear panel INPUT jack.
4. Press ENTER to complete offset voltage calibration.
5. Press EXIT to return to normal display.

Table of Contents

Other manuals for Keithley 6487

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6487 and is the answer not in the manual?

Keithley 6487 Specifications

General IconGeneral
TypePicoammeter/Voltage Source
Current Measurement Range20 fA to 20 mA
InterfacesGPIB, RS-232
Voltage Burden< 200 µV
Input Impedance>10^14 Ω
Power100 V / 120 V / 220 V / 240 V
Basic Current Measurement Accuracy0.4%

Related product manuals