Model 6487 Reference Manual Remote Operation 9-3
Interface selection and configuration procedures
klqb The unit will reset if the language is changed (SCPI, 488.1, and DDC).
When you select (enable) the GPIB interface, the RS-232 interface disables. Conversely,
selecting (enabling) the RS-232 interface disables the GPIB interface.
Select the interface from the COMM menu structure (access by pressing the COMM key
while in local). Use the RANGE keys to change the selected interface (RS-232 or GPIB).
Press ENTER to save the change.
klqb When an interface is enabled (on) or disabled (off), the instrument will exit from
the menu structure and perform the power-on sequence.
Configuring the GPIB interface
Select the GPIB interface from the COMM menu structure (access by pressing the
COMM key while in local). After selecting the GPIB interface, press the CONFIG key and
then the COMM key to configure the GPIB address and language. From this menu you can
check or change the following settings:
Primary address: 0–30
Language: SCPI (SCPI 1996.0 which includes 488.2), 488.1, or DDC
Press the RANGE keys to scroll through ADDress and LANGuage available. To make
changes to a setting, press the cursor right key. Then use the RANGE keys and the cursor
keys to select and modify the value as desired. Press ENTER to save the changes and stay
in the menu (pressing EXIT also saves the changes but leaves the menu).
RS-232 interface
klqb Only the SCPI language can be used with the RS-232 interface. The instrument
defaults to the SCPI language when the RS-232 interface is selected (enabled).
Select the RS-232 interface from the COMM menu structure (access by pressing the
COMM key while in local). After selecting the RS-232 interface, press the CONFIG key
and then the COMM key to configure the RS-232 interface and check or change the fol-
lowing settings:
BAUD: Baud rate (57.6K, 38.4k, 19.2k, 9600, 4800, 2400, 1200, 600, or 300)
BITS: Data bits (7 or 8)
PARITY: Parity (NONE, ODD, or EVEN)
TX TERMINATOR: Terminator (CR, LF, CRLF, or LFCR)
FLOW: Flow control (NONE or Xon/Xoff)
klqb See “RS-232 interface reference,” on page 9-16 for information on RS-232
settings and connections to the computer.

Table of Contents

Other manuals for Keithley 6487

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6487 and is the answer not in the manual?

Keithley 6487 Specifications

General IconGeneral
TypePicoammeter/Voltage Source
Current Measurement Range20 fA to 20 mA
InterfacesGPIB, RS-232
Voltage Burden< 200 µV
Input Impedance>10^14 Ω
Power100 V / 120 V / 220 V / 240 V
Basic Current Measurement Accuracy0.4%