EasyManua.ls Logo

Keithley 7001

Keithley 7001
353 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IEEE-488 Reference
5-134
Power-up Saved power-on setup
*RST 1 (all layers)
:SYSTem:PRESet 1 (all layers)
:ILINe? Query programmed input line #
Short-form format: :arm:tcon:asyn:ilin?
:arm:lay2:tcon:asyn:ilin?
:trig:tcon:asyn:ilin?
Response message: 1 to 6
These commands are used to select an input line for the asynchronous Trigger Link of the spec-
ified layer. Keep in mind that asynchronous Trigger Link input and output (see :OLINe) cannot
share the same line. If you assign the input to a line that is already being used for the output, an
error will occur and the command will be ignored.
10 OUTPUT 707; “:trig:tcon:asyn:ilin 3; ilin?”
20 ENTER 707; A$
30 PRINT A$
40 END
Line 10 Assigns the asynchronous Trigger Link input of the Trigger Layer to line #3, and then
queries the programmed input line.
Line 20 Addresses the 7001 to talk.
Line 30 Displays the programmed input line # (3).
:ASYNchronous:OLINe <NRf>
:ARM[:SEQuence[1]][:LAYer[1]]:TCONfigure:ASYNchronous:OLINe <NRf> Select output trigger line; Arm Layer 1
:ARM[:SEQuence[1]]:LAYer2:TCONfigure:ASYNchronous:OLINe <NRf> Select output trigger line; Arm Layer 2
:TRIGger[:SEQuence[1]]:TCONfigure:ASYNchronous:OLINe <NRf> Select output trigger line; Trigger Layer
<NRf> = 1 Line #1
= 2 Line #2
= 3 Line #3
= 4 Line #4
= 5 Line #5
= 6 Line #6
:arm:tcon:asyn:olin <NRf>
:arm:lay2:tcon:asyn:olin <NRf>
:trig:tcon:asyn:olin <NRf>
Defaults
Query
commands
Description
Programming
example
Parameters
Short-form
formats
Trigger subsystem
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Other manuals for Keithley 7001