IEEE-488 Reference
5-134
Power-up Saved power-on setup
*RST 1 (all layers)
:SYSTem:PRESet 1 (all layers)
:ILINe? Query programmed input line #
Short-form format: :arm:tcon:asyn:ilin?
:arm:lay2:tcon:asyn:ilin?
:trig:tcon:asyn:ilin?
Response message: 1 to 6
These commands are used to select an input line for the asynchronous Trigger Link of the spec-
ified layer. Keep in mind that asynchronous Trigger Link input and output (see :OLINe) cannot
share the same line. If you assign the input to a line that is already being used for the output, an
error will occur and the command will be ignored.
10 OUTPUT 707; “:trig:tcon:asyn:ilin 3; ilin?”
20 ENTER 707; A$
30 PRINT A$
40 END
Line 10 Assigns the asynchronous Trigger Link input of the Trigger Layer to line #3, and then
queries the programmed input line.
Line 20 Addresses the 7001 to talk.
Line 30 Displays the programmed input line # (3).
:ASYNchronous:OLINe <NRf>
:ARM[:SEQuence[1]][:LAYer[1]]:TCONfigure:ASYNchronous:OLINe <NRf> Select output trigger line; Arm Layer 1
:ARM[:SEQuence[1]]:LAYer2:TCONfigure:ASYNchronous:OLINe <NRf> Select output trigger line; Arm Layer 2
:TRIGger[:SEQuence[1]]:TCONfigure:ASYNchronous:OLINe <NRf> Select output trigger line; Trigger Layer
<NRf> = 1 Line #1
= 2 Line #2
= 3 Line #3
= 4 Line #4
= 5 Line #5
= 6 Line #6
:arm:tcon:asyn:olin <NRf>
:arm:lay2:tcon:asyn:olin <NRf>
:trig:tcon:asyn:olin <NRf>
Defaults
Query
commands
Description
Programming
example
Parameters
Short-form
formats
Trigger subsystem
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com