IEEE-488 Reference
5-135
Power-up Saved power-on setup
*RST 2 (all layers)
:SYSTem:PRESet 2 (all layers)
:OLINe? Query programmed output line #
Short-form formats: :arm:tcon:asyn:olin?
:arm:lay2:tcon:asyn:olin?
:trig:tcon:asyn:olin?
Response message: 1 to 6
These commands are used to select an output line for the asynchronous Trigger Link of the spec-
ified layer. Keep in mind that asynchronous Trigger Link input and output cannot share the same
line. If you assign the output to a line that is already being used for the input, an error will occur
and the command will be ignored.
10 OUTPUT 707; “:trig:tcon:asyn:olin 4; olin?”
20 ENTER 707; A$
30 PRINT A$
40 END
Line 10 Assigns the asynchronous Trigger Link output of the Trigger Layer to line #4, and
then queries the programmed output line.
Line 20 Addresses the 7001 to talk.
Line 30 Displays the programmed output line # (4).
:SSYNchronous:LINE <NRf>
:TRIGger[:SEQuence[1]]:TCONfigure:SSYNchronous Specify semi-synchronous Trigger Link line for the Trigger
Layer
<NRf> = 1 Line #1
= 2 Line #2
= 3 Line #3
= 4 Line #4
= 5 Line #5
= 6 Line #6
:trig:tcon:ssyn:line <NRf>
Power-up Saved power-on setup
*RST 1 (Trigger Layer)
:SYSTem:PRESet 1 (Trigger Layer)
Defaults
Query
commands
Description
Programming
example
Parameters
Short-form
format
Defaults
Trigger subsystem